Back to Search
Start Over
Bilateral comparison of C activity measurements at the NCBJ RC POLATOM and the ENEA-INMRI.
- Source :
-
Journal of Radioanalytical & Nuclear Chemistry . Nov2017, Vol. 314 Issue 2, p721-725. 5p. 4 Charts, 1 Graph. - Publication Year :
- 2017
-
Abstract
- A bilateral comparison of C activity measurements between ENEA-INMRI (Italy) and NCBJ RC POLATOM (Poland) was organized in the year 2015 and piloted by POLATOM, which provided the sources for the comparison. A set of six sources in Ultima Gold scintillator was measured at POLATOM using four different counters by two liquid scintillation counting methods: TDCR and CIEMAT/NIST. Three of sources were measured also at ENEA-INMRI by using two different TDCR counters. The results achieved in the comparison show an excellent agreement. [ABSTRACT FROM AUTHOR]
- Subjects :
- *SCINTILLATORS
*LUMINESCENCE
*PARTICLE detectors
*SCINTILLATION counters
*PHOSPHORS
Subjects
Details
- Language :
- English
- ISSN :
- 02365731
- Volume :
- 314
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- Journal of Radioanalytical & Nuclear Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 125894383
- Full Text :
- https://doi.org/10.1007/s10967-017-5464-y