Digital gamma-gamma coincidence HPGe system for environmental analysis.

Authors :
Marković, Nikola
Roos, Per
Nielsen, Sven Poul
Source :
Applied Radiation & Isotopes. Aug2017, Vol. 126, p194-196. 3p.
Publication Year :
2017

Abstract

The performance of a new gamma-gamma coincidence spectrometer system for environmental samples analysis at the Center for Nuclear Technologies of the Technical University of Denmark (DTU) is reported. Nutech Coincidence Low Energy Germanium Sandwich (NUCLeGeS) system consists of two HPGe detectors in a surface laboratory with a digital acquisition system used to collect the data in time-stamped list mode with 10 ns time resolution. The spectrometer is used in both anticoincidence and coincidence modes. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09698043
Volume :
126
Database :
Academic Search Index
Journal :
Applied Radiation & Isotopes
Publication Type :
Academic Journal
Accession number :
123658882
Full Text :
https://doi.org/10.1016/j.apradiso.2016.12.017