Abstract

Quantification and mapping possibilities of lithium in geological material, by Nuclear Reaction Analysis (NRA), was evaluated at the Lund Ion Beam Analysis Facility (LIBAF). LiF and two Standard Reference Materials, (SRM 610 and SRM 612) were used in the investigation. The main part of the data was obtained at the beam energy 635 keV studying the high Q-value reaction 7 Li(p, α ) 4 He, but reaction yield and detection limits were also briefly investigated as a function of the energy. A double-sided silicon strip detector (DSSSD) was used to detect the α -particles emitted in the reaction in the backward direction. The combination of the high Q-value, a reasonably good cross-section and the possibility to use a high beam current have been demonstrated to allow for measurement of concentrations down below 50 ppm. Proton energies below 800 keV were demonstrated to be appropriate energies for extracting lithium in combination with boron analysis. [ABSTRACT FROM AUTHOR]