Validation of image cytometry for sperm concentration measurement: Comparison with manual counting of 4010 human semen samples.

Authors :
Egeberg Palme, Dorte L.
Johannsen, Trine Holm
Petersen, Jørgen Holm
Skakkebæk, Niels E.
Juul, Anders
Jørgensen, Niels
Almstrup, Kristian
Source :
Clinica Chimica Acta. May2017, Vol. 468, p114-119. 6p.
Publication Year :
2017

Abstract

Sperm concentration is an essential parameter in the diagnostic evaluation of men from infertile couples. It is usually determined by manual counting using a hemocytometer, and is therefore both laborious and subjective. We have earlier shown that a newly developed image cytometry (IC) method may be used to determine sperm concentration. Here we present a validation of the IC method by analysis of 4010 semen samples. There was high agreement between IC and manual counting at sperm concentrations above 3 mill/ml and in samples with concentrations above 12 mill/ml the two methods can be used interchangeable. However, we found substantial differences in samples below 3 mill/ml. We also assessed the accuracy of the two methods by repeated measurements of 248 samples, which revealed that IC measurements seemed more accurate. Moreover, based on ten samples counted by several operators the IC method had a lower coefficient of variation than the manual method (5% vs 10%), indicating a better precision of the IC method. In conclusion, measurement of sperm concentration by IC can be used at concentrations above 3 mill/ml and seems more accurate and precise than manual counting, making it an attractive option in the daily clinical practice. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00098981
Volume :
468
Database :
Academic Search Index
Journal :
Clinica Chimica Acta
Publication Type :
Academic Journal
Accession number :
123157402
Full Text :
https://doi.org/10.1016/j.cca.2017.02.014