Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene.
- Source :
-
PLoS ONE . 1/20/2017, Vol. 12 Issue 1, p1-10. 10p. - Publication Year :
- 2017
-
Abstract
- More than 60.000 hip arthroplasty are performed every year in Italy. Although Ultra-High-Molecular-Weight-Polyethylene remains the most used material as acetabular cup, wear of this material induces over time in vivo a foreign-body response and consequently osteolysis, pain, and the need of implant revision. Furthermore, oxidative wear of the polyethylene provoke several and severe failures. To solve these problems, highly cross-linked polyethylene and Vitamin-E-stabilized polyethylene were introduced in the last years. In in vitro experiments, various efforts have been made to compare the wear behavior of standard PE and vitamin-E infused liners. In this study we compared the in vitro wear behavior of two different configurations of cross-linked polyethylene (with and without the add of Vitamin E) vs. the standard polyethylene acetabular cups. The aim of the present study was to validate a micro X-ray computed tomography technique to assess the wear of different commercially available, polyethylene’s acetabular cups after wear simulation; in particular, the gravimetric method was used to provide reference wear values. The agreement between the two methods is documented in this paper. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 19326203
- Volume :
- 12
- Issue :
- 1
- Database :
- Academic Search Index
- Journal :
- PLoS ONE
- Publication Type :
- Academic Journal
- Accession number :
- 120873898
- Full Text :
- https://doi.org/10.1371/journal.pone.0170263