A new method to calibrate the absolute sensitivity of a soft X-ray streak camera.

Authors :
Jian Yu
Shenye Liu
Jin Li
Zhiwen Yang
Ming Chen
Luting Guo
Li Yao
Shali Xiao
Source :
Review of Scientific Instruments. Dec2016, Vol. 87 Issue 12, p1-5. 5p. 4 Diagrams, 3 Graphs.
Publication Year :
2016

Abstract

In this paper, we introduce a new method to calibrate the absolute sensitivity of a soft X-ray streak camera (SXRSC). The calibrations are done in the static mode by using a small laser-produced X-ray source. A calibrated X-ray CCD is used as a secondary standard detector to monitor the X-ray source intensity. In addition, two sets of holographic flat-field grating spectrometers are chosen as the spectral discrimination systems of the SXRSC and the X-ray CCD. The absolute sensitivity of the SXRSC is obtained by comparing the signal counts of the SXRSC to the output counts of the X-ray CCD. Results show that the calibrated spectrum covers the range from 200 eV to 1040 eV. The change of the absolute sensitivity in the vicinity of the K-edge of the carbon can also be clearly seen. The experimental values agree with the calculated values to within 29% error. Compared with previous calibration methods, the proposed method has several advantages: a wide spectral range, high accuracy, and simple data processing. Our calibration results can be used to make quantitative X-ray flux measurements in laser fusion research. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
87
Issue :
12
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
120465470
Full Text :
https://doi.org/10.1063/1.4971847