Relative calibration of energy thresholds on multi-bin spectral x-ray detectors.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Dec2016, Vol. 840, p1-4. 4p. - Publication Year :
- 2016
-
Abstract
- Accurate and reliable energy calibration of spectral x-ray detectors used in medical imaging is essential for avoiding ring artifacts in the reconstructed images (computed tomography) and for performing accurate material basis decomposition. A simple and accurate method for relative calibration of the energy thresholds on a multi-bin spectral x-ray detector is presented. The method obtains the linear relations between all energy thresholds in a channel by scanning the thresholds with respect to each other during x-ray illumination. The method does not rely on a model of the detector's response function and does not require any identifiable features in the x-ray spectrum. Applying the same method, the offset between the thresholds can be determined also without external stimuli by utilizing the electronic noise as a source. The simplicity and accuracy of the method makes it suitable for implementation in clinical multi-bin spectral x-ray imaging systems. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 840
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 120016487
- Full Text :
- https://doi.org/10.1016/j.nima.2016.09.045