Back to Search Start Over

Evaluation of the detection features of UD-807 dosimeters and their suitability for use in personal dosimetry.

Authors :
Skubacz, Krystian
Source :
Radiation Measurements. Dec2016, Vol. 95, p16-24. 9p.
Publication Year :
2016

Abstract

This paper contains the results of studies into the possibilities and limitations of the use of UD-807 Panasonic dosimeters in extremity dosimetry. Detection efficiency, critical levels, detection limits, energy and angular dependence were determined with respect to the beta radiation emitted by 90 Sr| 90 Y and photon radiation in the range of 33 keV–662 keV. The calibration coefficients, expressed in units of personal dose equivalent H p (0.07), differed by no more than 10% for all types of radiation when the radiation reached the dosimeters at an angle of 0°. In other configurations, if the angle between the normal to the dosimeter surface and the radiation direction increased, the detection efficiency rapidly decreased - to a lesser extent in relation to the 33 keV photon radiation than to the beta radiation. The reduction was insignificant for photon radiation emitted by 137 Cs. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
13504487
Volume :
95
Database :
Academic Search Index
Journal :
Radiation Measurements
Publication Type :
Academic Journal
Accession number :
119781321
Full Text :
https://doi.org/10.1016/j.radmeas.2016.11.001