Back to Search
Start Over
Evaluation of the detection features of UD-807 dosimeters and their suitability for use in personal dosimetry.
- Source :
-
Radiation Measurements . Dec2016, Vol. 95, p16-24. 9p. - Publication Year :
- 2016
-
Abstract
- This paper contains the results of studies into the possibilities and limitations of the use of UD-807 Panasonic dosimeters in extremity dosimetry. Detection efficiency, critical levels, detection limits, energy and angular dependence were determined with respect to the beta radiation emitted by 90 Sr| 90 Y and photon radiation in the range of 33 keV–662 keV. The calibration coefficients, expressed in units of personal dose equivalent H p (0.07), differed by no more than 10% for all types of radiation when the radiation reached the dosimeters at an angle of 0°. In other configurations, if the angle between the normal to the dosimeter surface and the radiation direction increased, the detection efficiency rapidly decreased - to a lesser extent in relation to the 33 keV photon radiation than to the beta radiation. The reduction was insignificant for photon radiation emitted by 137 Cs. [ABSTRACT FROM AUTHOR]
- Subjects :
- *DOSIMETERS
*RADIATION dosimetry
*DETECTION limit
*PHOTON emission
*BETA rays
Subjects
Details
- Language :
- English
- ISSN :
- 13504487
- Volume :
- 95
- Database :
- Academic Search Index
- Journal :
- Radiation Measurements
- Publication Type :
- Academic Journal
- Accession number :
- 119781321
- Full Text :
- https://doi.org/10.1016/j.radmeas.2016.11.001