Left ventricular function assessment using I/Tc dual-isotope acquisition with two semi-conductor cadmium-zinc-telluride (CZT) cameras: a gated cardiac phantom study.

Authors :
Blaire, Tanguy
Bailliez, Alban
Bouallegue, Fayçal
Bellevre, Dimitri
Agostini, Denis
Manrique, Alain
Source :
EJNMMI Physics. 11/11/2016, Vol. 3 Issue 1, p1-11. 11p.
Publication Year :
2016

Abstract

Background: The impact of increased energy resolution of cadmium-zinc-telluride (CZT) cameras on the assessment of left ventricular function under dual-isotope conditions (Tc and I) remains unknown. The Amsterdam-gated dynamic cardiac phantom (AGATE, Vanderwilt techniques, Boxtel, The Netherlands) was successively filled with a solution of I alone, Tc alone, and a mixture of I and Tc. A total of 12 datasets was acquired with each commercially available CZT camera (DNM 530c, GE Healthcare and DSPECT, Biosensors International) using both energy windows (Tc or I) with ejection fraction set to 33, 45, and 60 %. End-diastolic (EDV) and end-systolic (ESV) volumes, ejection fraction (LVEF), and regional wall motion and thickening (17-segment model) were assessed using Cedars-Sinai QGS Software. Concordance between single- and dual-isotope acquisitions was tested using Lin's concordance correlation coefficient (CCC) and Bland-Altman plots. Results: There was no significant difference between single- or simultaneous dual-isotope acquisition (I and Tc) for EDV, ESV, LVEF, or segmental wall motion and thickening. Myocardial volumes using single- (I, Tc) and dual-isotope (reconstructed using both I and Tc energy windows) acquisitions were, respectively, the following: EDV (mL) 88 ± 27 vs. 89 ± 27 vs. 92 ± 29 vs. 90 ± 26 for DNM 530c ( p = NS) and 82 ± 20 vs. 83 ± 22 vs. 79 ± 19 vs. 77 ± 20 for DSPECT ( p = NS); ESV (mL) 40 ± 1 vs. 41 ± 2 vs. 41 ± 2 vs. 42 ± 1 for DNM 530c ( p = NS) and 37 ± 5 vs. 37 ± 1 vs. 35 ± 3 vs. 34 ± 2 for DSPECT ( p = NS); LVEF (%) 52 ± 14 vs. 51 ± 13 vs. 53 ± 13 vs. 51 ± 13 for DNM 530c ( p = NS) and 52 ± 16 vs. 54 ± 13 vs. 54 ± 14 vs. 54 ± 13 for DSPECT ( p = NS); regional motion (mm) 6.72 ± 2.82 vs. 6.58 ± 2.52 vs. 6.86 ± 2.99 vs. 6.59 ± 2.76 for DNM 530c ( p = NS) and 6.79 ± 3.17 vs. 6.81 ± 2.75 vs. 6.71 ± 2.50 vs. 6.62 ± 2.74 for DSPECT ( p = NS). The type of camera significantly impacted only on ESV ( p < 0.001). Conclusions: The new CZT cameras yielded similar results for the assessment of LVEF and regional motion using different energy windows (I or Tc) and acquisition types (single vs. dual). With simultaneous dual-isotope acquisitions, the presence of I did not impact on LVEF assessment within the Tc energy window for either CZT camera. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
21977364
Volume :
3
Issue :
1
Database :
Academic Search Index
Journal :
EJNMMI Physics
Publication Type :
Academic Journal
Accession number :
119435411
Full Text :
https://doi.org/10.1186/s40658-016-0163-2