Back to Search
Start Over
Quantitative aspects of inductively coupled plasma mass spectrometry.
- Source :
-
Philosophical Transactions of the Royal Society A: Mathematical, Physical & Engineering Sciences . 10/28/2016, Vol. 374 Issue 2079, p1-18. 18p. - Publication Year :
- 2016
-
Abstract
- Accurate determination of elements in various kinds of samples is essential for many areas, including environmental science, medicine, as well as industry. Inductively coupled plasma mass spectrometry (ICP-MS) is a powerful tool enabling multi-elemental analysis of numerous matrices with high sensitivity and good precision. Various calibration approaches can be used to perform accurate quantitative measurements by ICP-MS. They include the use of pure standards, matrix-matched standards, or relevant certified reference materials, assuring traceability of the reported results. This review critically evaluates the advantages and limitations of different calibration approaches, which are used in quantitative analyses by ICP-MS. Examples of such analyses are provided. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 1364503X
- Volume :
- 374
- Issue :
- 2079
- Database :
- Academic Search Index
- Journal :
- Philosophical Transactions of the Royal Society A: Mathematical, Physical & Engineering Sciences
- Publication Type :
- Academic Journal
- Accession number :
- 118896910
- Full Text :
- https://doi.org/10.1098/rsta.2015.0369