A multi-wavelength, high-contrast contact radiography system for the study of low-density aerogel foams.

Authors :
Opachich, Y. P.
Koch, J. A.
Haugh, M. J.
Romano, E.
Lee, J. J.
Huffman, E.
Weber, F. A.
Bowers, J. W.
Benedetti, L. R.
Wilson, M.
Prisbrey, S. T.
Wehrenberg, C. E.
Baumann, T. F.
Lenhardt, J. M.
Cook, A.
Arsenlis, A.
Park, H.-S.
Remington, B. A.
Source :
Review of Scientific Instruments. Jul2016, Vol. 87 Issue 7, p073706-1-073706-5. 5p. 1 Diagram, 4 Charts, 5 Graphs.
Publication Year :
2016

Abstract

A multi-wavelength, high contrast contact radiography system has been developed to characterize density variations in ultra-low density aerogel foams. These foams are used to generate a ramped pressure drive in materials strength experiments at the National Ignition Facility and require precision characterization in order to reduce errors in measurements. The system was used to characterize density variations in carbon and silicon based aerogels to ∼10.3% accuracy with ∼30 μm spatial resolution. The system description, performance, and measurement results collected using a 17.8 mg/cc carbon based JX-6 (C20H30) aerogel are discussed in this manuscript. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
87
Issue :
7
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
117132106
Full Text :
https://doi.org/10.1063/1.4958826