Back to Search Start Over

Modeling single cell antibody excretion on a biosensor.

Authors :
Stojanović, Ivan
Baumgartner, Wolfgang
van der Velden, Thomas J.G.
Terstappen, Leon W.M.M.
Schasfoort, Richard B.M.
Source :
Analytical Biochemistry. Jul2016, Vol. 504, p1-3. 3p.
Publication Year :
2016

Abstract

We simulated, using Comsol Multiphysics, the excretion of antibodies by single hybridoma cells and their subsequent binding on a surface plasmon resonance imaging (SPRi) sensor. The purpose was to confirm that SPRi is suitable to accurately quantify antibody (anti-EpCAM) excretion. The model showed that antibody loss by diffusion away from the sensor was less than 1%. Unexpectedly, more than 99% of the excreted antibodies were captured on the sensor. These data prove the remarkable phenomenon that the SPRi output of cellular antibody excretion and its subsequent binding, performed under the conditions described here, is directly usable for quantification of single cell antibody production rates. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00032697
Volume :
504
Database :
Academic Search Index
Journal :
Analytical Biochemistry
Publication Type :
Academic Journal
Accession number :
115492279
Full Text :
https://doi.org/10.1016/j.ab.2016.03.018