Back to Search Start Over

Channeltron operating in coincidence mode

Authors :
Krečak, Z.
Krčmar, M.
Bogovac, M.
Ljubičić, A.
Source :
Nuclear Instruments & Methods in Physics Research Section B. Jan2004, Vol. 213 Issue 1-4, p310. 5p.
Publication Year :
2004

Abstract

We have investigated channeltron operating in coincidence with silicon surface barrier detector for detection of extreme low energy particles. By use of Po-209 radionuclide and silver backing, a radioactive source was prepared as a fine emitter of released α particles, conversion electrons, recoil Pb-205 ions and knock out silver ions. To avoid parasite coincidence events mainly from channeltron anode RC circuit, we have applied appropriate Faraday screens. After getting proper counting rates, composite spectra were recorded. Based on time-of-flight and secondary electron emission coefficient arguments, particular spectra of mentioned particles were extracted. Final experimental data analysis enabled us to found channeltron efficiency curves for electrons, lead ions and silver ions with low detection thresholds at 40 eV, 4 keV and 1.7 keV, respectively. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
0168583X
Volume :
213
Issue :
1-4
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section B
Publication Type :
Academic Journal
Accession number :
11471830
Full Text :
https://doi.org/10.1016/S0168-583X(03)01634-3