Back to Search
Start Over
Monitoring multicomponent quality traits in tomato juice using portable mid-infrared (MIR) spectroscopy and multivariate analysis.
- Source :
-
Food Control . Aug2016, Vol. 66, p79-86. 8p. - Publication Year :
- 2016
-
Abstract
- In this study, a novel portable infrared spectrometer was evaluated for rapid and simultaneous determination of glucose, fructose, total reducing sugars, soluble solids (°Brix), serum viscosity, Bostwick consistency, predicted paste Bostwick value and pH. A total of 350 hot-break juices from two consecutive years (2013 and 2014) including 66 different varieties grown in 6 counties of California, USA, were used in this study. Duplicate spectra for each tomato juice were collected using the transmission accessory of the portable infrared spectrometer at 50 μm fixed path length. Samples were randomly divided into calibration (n = 280) and external validation (n = 70) sets and partial least squares (PLS) regression was used to develop calibration models to predict all the variables (quality traits) in the validation samples based on the tomato juice spectral data (r pred > 0.82). Overall, our validated chemometric models allowed rapidly (∼2 min) predicting all the quality traits in tomato juice samples with no sample preparation. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09567135
- Volume :
- 66
- Database :
- Academic Search Index
- Journal :
- Food Control
- Publication Type :
- Academic Journal
- Accession number :
- 114313977
- Full Text :
- https://doi.org/10.1016/j.foodcont.2016.01.031