YAP-(S)PET Small Animal Scanner: Quantitative Results.

Authors :
Di Domenico, G.
Zavattini, G.
Moretti, E.
Piffanelli, A.
Giganti, M.
Motta, A.
Sabba, N.
Uccelli, L.
Benini, E.
Duatti, A.
Bolzati, C.
Boschi, A.
Del Guerra, A.
Source :
IEEE Transactions on Nuclear Science. Oct2003 Part 2 of 3, Vol. 50 Issue 5, p1351. 6p.
Publication Year :
2003

Abstract

The University of Ferrara YAP-(S)PET scanner is currently being employed in small animal SPECT studies, with [sup99m]Tc labeled radiotracers, with the goal of obtaining quantitative activity measurements from region of interest (ROI) analysis of reconstructed images of rats. The measurements will be compared directly with traditional ex vivo measurement of activity with gamma counters. To achieve this goal, the scanner was calibrated relative to a standard Isodose dose calibrator which was calibrated with various certified activity sources. The calibration factor K was defined as the ratio of the activity measured with the Isodose (MBq) and the image count rate (cps). We find K = 0.1346 ± 0.0008 MBq/cps, with good linearity over a wide range of activities (9-88 MBq). With the calibrated scanner we compared results of activity measurements from images of whole-rat heart acquisitions versus excised hearts. We found good agreement between the two measurements. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189499
Volume :
50
Issue :
5
Database :
Academic Search Index
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
11422622
Full Text :
https://doi.org/10.1109/TNS.2003.817970