Back to Search
Start Over
Characteristics of GRIFFIN high-purity germanium clover detectors.
- Source :
-
Nuclear Instruments & Methods in Physics Research Section A . Jun2016, Vol. 820, p126-131. 6p. - Publication Year :
- 2016
-
Abstract
- The Gamma-Ray Infrastructure For Fundamental Investigations of Nuclei, GRIFFIN, is a new experimental facility for radioactive decay studies at the TRIUMF-ISAC laboratory. The performance of the 16 high-purity germanium (HPGe) clover detectors that will make up the GRIFFIN spectrometer is reported. The energy resolution, efficiency, timing resolution, crosstalk and preamplifier properties of each crystal were measured using a combination of analog and digital data acquisition techniques. The absolute efficiency and add-back factors are determined for the energy range of 80–3450 keV. The detectors show excellent performance with an average over all 64 crystals of a FWHM energy resolution of 1.89(6) keV and relative efficiency with respect to a 3 in . × 3 in . NaI detector of 41(1)% at 1.3 MeV. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 01689002
- Volume :
- 820
- Database :
- Academic Search Index
- Journal :
- Nuclear Instruments & Methods in Physics Research Section A
- Publication Type :
- Academic Journal
- Accession number :
- 114129064
- Full Text :
- https://doi.org/10.1016/j.nima.2016.03.016