Back to Search
Start Over
Stability evaluation and correction of a pulsed neutron generator prompt gamma activation analysis system.
- Source :
-
Journal of Radioanalytical & Nuclear Chemistry . Mar2016, Vol. 307 Issue 3, p1781-1786. 6p. 1 Diagram, 2 Charts, 2 Graphs. - Publication Year :
- 2016
-
Abstract
- Source output stability is important for accurate measurement in prompt gamma neutron activation. This is especially true when measuring low-concentration elements such as in vivo nitrogen (~2.5 % of body weight). We evaluated the stability of the compact DT neutron generator within an in vivo nitrogen measurement system. Review of gamma event/time patterns and data from an auxiliary detector showed significant variations among repeated phantom runs. Neutron generator instability had a significant effect on measurement precision. The neutron generator used in our system must be monitored for output consistency. Adjustments must be made to measurement results to correct for generator instability. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 02365731
- Volume :
- 307
- Issue :
- 3
- Database :
- Academic Search Index
- Journal :
- Journal of Radioanalytical & Nuclear Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 113529212
- Full Text :
- https://doi.org/10.1007/s10967-015-4440-7