Back to Search Start Over

Performance Evaluation of a Fluid Temperature-Compensated Single Sensor Constant Temperature Anemometer.

Authors :
Ferreira, Rômulo Pires Coelho
Freire, Raimundo Carlos Silvério
Deep, Gurdip Singh
Source :
IEEE Transactions on Instrumentation & Measurement. Oct2003, Vol. 52 Issue 5, p1554-1558. 5p.
Publication Year :
2003

Abstract

The performance of a hot-wire anemometer configuration is affected by variation in the fluid temperature The classical temperature compensation techniques in such anemometers employ two sensors. The performance of a temperature-compensated hot-wire anemometer configuration using a single sensor alternating between two Operating temperature and proposed for constant fluid velocity is investigated under conditions of time-varying fluid velocity. The measurement error introduced is quantified and can be practically eliminated using low-pass digital filter. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
52
Issue :
5
Database :
Academic Search Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
11331508
Full Text :
https://doi.org/10.1109/TIM.2003.818554