Effect of increased load on the wear of a large diameter metal-on-metal modular hip prosthesis with a high inclination angle of the acetabular cup.

Authors :
Saikko, Vesa
Source :
Tribology International. Apr2016, Vol. 96, p149-154. 6p.
Publication Year :
2016

Abstract

In some earlier hip simulator studies of large-diameter metal-on-metal (MoM) hip prostheses high, unexplained variation in wear has been observed. In the present study it was found that high variation occurs when the tribological endurance limit of the MoM device in question is being approached. When the limit is exceeded, high wear takes place uniformly, and the variation is low. Below the limit, both the wear and its variation are low. The endurance limit was exceeded by increasing the peak load from 2 kN to 3 kN, to simulate obesity. The acetabular cup inclination angle was high, 70°. Potential clinical wear problems can be predicted by hip simulator studies provided that relevant adverse test conditions are included. [ABSTRACT FROM AUTHOR]