Back to Search Start Over

Fast scintillation timing detector using proportional-mode avalanche photodiode for nuclear resonant scattering experiments in high-energy synchrotron X-ray region.

Authors :
Inoue, Keisuke
Kishimoto, Shunji
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jan2016, Vol. 806, p420-424. 5p.
Publication Year :
2016

Abstract

To obtain both a high count rate of >10 7 s −1 and a detection efficiency sufficient for high-energy X-rays of >30 keV, we propose a scintillation timing detector using a proportional-mode silicon avalanche photodiode (Si-APD) for synchrotron radiation nuclear resonant scattering. We here present results obtained with a prototype detector using a lead-loaded plastic scintillator (EJ-256) mounted on a proportional-mode Si-APD (active area size: 3 mm in diameter). The detector was operated at ‒35 °C for a better signal-to-noise ratio. Using synchrotron X-rays of 67.41 keV, which is the same energy as the first excited level of 61 Ni, we successfully measured pulse-height and time spectra of the scintillation light. A good time resolution of 0.50±0.06 ns (full width at half-maximum) was obtained for 67.41 keV X-rays with a scintillator 3 mm in diameter and 2 mm thick. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
806
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
110942318
Full Text :
https://doi.org/10.1016/j.nima.2015.10.035