Back to Search
Start Over
Sensitive Determination of Amantadine in Microdialysis Samples from Rat Plasma by HPLC with Fluorescence Detection.
- Source :
-
Journal of Liquid Chromatography & Related Technologies . 2015, Vol. 38 Issue 17, p1622-1628. 7p. 1 Black and White Photograph, 4 Charts, 1 Graph. - Publication Year :
- 2015
-
Abstract
- A new and sensitive high-performance liquid chromatography (HPLC) method with 10-methyl-acridone-2-sulfonyl chloride (MASC) as pre-column fluorescence labeling reagent was developed for the determination of amantadine in microdialysis samples from rat plasma. The HPLC sensitivity of amantadine was greatly enhanced by introducing MASC with large conjugated structure and excellent fluorescence property into the amantadine molecule. The limit of detection was 6.8 ng mL−1and only 30 μL sample was needed for the pharmacokinetic study. The linearity range was 25–500 ng mL−1with correlation coefficients of 0.9989. The proposed method was successfully applied to the determination of amantadine in microdialysis samples from rat plasma. It also shows powerful potential for the pharmacokinetic study of compounds with amino group. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 10826076
- Volume :
- 38
- Issue :
- 17
- Database :
- Academic Search Index
- Journal :
- Journal of Liquid Chromatography & Related Technologies
- Publication Type :
- Academic Journal
- Accession number :
- 110567865
- Full Text :
- https://doi.org/10.1080/10826076.2015.1087026