Back to Search Start Over

A Battery Life beyond His 'Expectancy'.

Authors :
STRÖKER, ERWIN
PAPPAERT, GUDRUN
ROELANDT, BRIAN
BRUGADA, PEDRO
de ASMUNDIS, CARLO
Source :
Pacing & Clinical Electrophysiology. Oct2015, Vol. 38 Issue 10, p1228-1230. 3p.
Publication Year :
2015

Abstract

The article presents a case study of a 77-year-old man with an implanted Medtronic Viva Quad XT Cardiac Resynchronization Therapy Defibrillator (CRT-D) device. The patient underwent an actual device interrogation which showed a Power-on-Reset (PoR) activity leading to an electrical reset of the device. An aberrant behavior of the battery life indicator on Meditronic devices can be caused by PoR.

Details

Language :
English
ISSN :
01478389
Volume :
38
Issue :
10
Database :
Academic Search Index
Journal :
Pacing & Clinical Electrophysiology
Publication Type :
Academic Journal
Accession number :
110067600
Full Text :
https://doi.org/10.1111/pace.12622