Development and performance of a new prosthesis system using ultrasonic sensor for wrist movements: a preliminary study.

Authors :
Abd Razak, Nasrul Anuar
Abu Osman, Noor Azuan
Gholizadeh, Hossein
Ali, Sadeeq
Source :
BioMedical Engineering OnLine. 2014, Vol. 13, p49-49. 1p.
Publication Year :
2014

Abstract

<bold>Background: </bold>The design and performance of a new development prosthesis system known as biomechatronics wrist prosthesis is presented in this paper. The prosthesis system was implemented by replacing the Bowden tension cable of body powered prosthesis system using two ultrasonic sensors, two servo motors and microcontroller inside the prosthesis hand for transradial user.<bold>Methods: </bold>The system components and hand prototypes involve the anthropometry, CAD design and prototyping, biomechatronics engineering together with the prosthetics. The modeler construction of the system develop allows the ultrasonic sensors that are placed on the shoulder to generate the wrist movement of the prosthesis. The kinematics of wrist movement, which are the pronation/supination and flexion/extension were tested using the motion analysis and general motion of human hand were compared. The study also evaluated the require degree of detection for the input of the ultrasonic sensor to generate the wrist movements.<bold>Results: </bold>The values collected by the vicon motion analysis for biomechatronics prosthesis system were reliable to do the common tasks in daily life. The degree of the head needed to bend to give the full input wave was about 45°-55° of rotation or about 14 cm-16 cm. The biomechatronics wrist prosthesis gave higher degree of rotation to do the daily tasks but did not achieve the maximum degree of rotation.<bold>Conclusion: </bold>The new development of using sensor and actuator in generating the wrist movements will be interesting for used list in medicine, robotics technology, rehabilitations, prosthetics and orthotics. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
1475925X
Volume :
13
Database :
Academic Search Index
Journal :
BioMedical Engineering OnLine
Publication Type :
Academic Journal
Accession number :
107853832
Full Text :
https://doi.org/10.1186/1475-925X-13-49