Back to Search Start Over

Diffractometer for soft x-ray resonant magnetic scattering.

Authors :
Grabis, J.
Nefedov, A.
Zabel, H.
Source :
Review of Scientific Instruments. Sep2003, Vol. 74 Issue 9, p4048. 4p.
Publication Year :
2003

Abstract

We report on the design and construction of a new diffractometer for soft x-ray resonant magnetic scattering which has been built at the Ruhr-University Bochum. The ultrahigh vacuum-compatible diffractometer comprises a two-circle goniometer and works in horizontal scattering geometry. Rotation of the detector and sample is realized by two differentially pumped rotating platforms with motors and gears external to the vacuum. The sample environment consists of a closed-cycle He cryostat that allows heating up to 600 K and applying a magnetic field of ±2.5 kOe. All functions of the experimental setup have been extensively tested at the BESSY II Synchrotron (beamline PM3). The experimental results demonstrate the performance of the instrument. © 2003 American Institute of Physics. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
74
Issue :
9
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
10612658
Full Text :
https://doi.org/10.1063/1.1602932