Back to Search
Start Over
Neutron Spectrometry with EJ299-33 Plastic Scintillator for E_n = 10 \! - \! 100 ~\MeV.
- Source :
-
IEEE Transactions on Nuclear Science . 6/20/2015 Part 3, Vol. 62 Issue 3c, p1422-1428. 7p. - Publication Year :
- 2015
-
Abstract
- A neutron spectrometer based on a single EJ299-33 plastic scintillator (5~\cm diameter\times 5~\cm) is described. A digital implementation of pulse shape discrimination is used to separate events associated with neutrons from those associated with gamma-rays. Measurements made using a ns-pulsed neutron beam are used to produce response functions of the detector for neutrons over the energy range 10 – 100 MeV. These lineshapes are used to test the capability of the spectrometer to produce neutron energy spectra, via the unfolding of pulse height spectra, for use in neutron fields having any type of time structure. The results are compared with those from a similar spectrometer constructed from EJ301 liquid scintillator. [ABSTRACT FROM PUBLISHER]
Details
- Language :
- English
- ISSN :
- 00189499
- Volume :
- 62
- Issue :
- 3c
- Database :
- Academic Search Index
- Journal :
- IEEE Transactions on Nuclear Science
- Publication Type :
- Academic Journal
- Accession number :
- 103264817
- Full Text :
- https://doi.org/10.1109/TNS.2015.2430455