Back to Search Start Over

Experimental investigation on two-stage thermoelectric cooling system adopted in isoelectric focusing.

Authors :
Hu, H.M.
Ge, T.S.
Dai, Y.J.
Wang, R.Z.
Source :
International Journal of Refrigeration. May2015, Vol. 53, p1-12. 12p.
Publication Year :
2015

Abstract

Performance of temperature control is crucial to the operation of isoelectric focusing equipment (IEF). In this paper, two-stage thermoelectric cooling module (TEM) is proposed to be adopted in IEF to realize prompt and precise temperature control as well as low focusing temperature (T f ). Three different prototypes including HP + baffle , AL + baffle and HP + fin are developed to obtain optimal design. Experimental setups of these prototypes are built up to test their performance. Temperature distribution on cooling plate, COP and air temperature in bottom chamber with respect to different T f s are adopted as performance indices to evaluate performance of these prototypes. Experimental results show that aluminum plate with heat pipes, used as cooling plate, can improve its temperature uniformity. Moreover, fin-type heat sink with baffle can effectively dissipate heat on the hot side of TEM with little impact on the other parts of IEF. The T f of HP + baffle can be kept at 10 °C. And its COP can reach 2.0 under general working condition. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01407007
Volume :
53
Database :
Academic Search Index
Journal :
International Journal of Refrigeration
Publication Type :
Academic Journal
Accession number :
102696670
Full Text :
https://doi.org/10.1016/j.ijrefrig.2015.01.018