Fat to muscle ratio measurements with dual energy x-ray absorbtiometry.

Authors :
Chen, A.
Luo, J.
Wang, A.
Broadbent, C.
Zhong, J.
Dilmanian, F.A.
Zafonte, F.
Zhong, Z.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jul2015, Vol. 788, p24-28. 5p.
Publication Year :
2015

Abstract

Accurate measurement of the fat-to-muscle ratio in animal model is important for obesity research. An efficient way to measure the fat to muscle ratio in animal model using dual-energy absorptiometry is presented in this paper. A radioactive source exciting x-ray fluorescence from a target material is used to provide the two x-ray energies needed. The x-rays, after transmitting through the sample, are measured with an energy-sensitive Ge detector. Phantoms and specimens were measured. The results showed that the method was sensitive to the fat to muscle ratios with good linearity. A standard deviation of a few percent in the fat to muscle ratio could be observed with the x-ray dose of 0.001 mGy. [ABSTRACT FROM AUTHOR]