Back to Search Start Over

Development of a DAQ system for a plasma display panel-based X-ray detector (PXD).

Authors :
Lee, Hakjae
Jung, Young-Jun
Eom, Sangheum
Kang, Jungwon
Lee, Kisung
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jun2015, Vol. 784, p213-219. 7p.
Publication Year :
2015

Abstract

Recently, a novel plasma display panel (PDP)-based X-ray detector (PXD) was developed. The goal of this study is to develop a data acquisition system for use with the PXD as an imaging detector. Since the prototype detector does not have any barrier ribs or a switching device in a detector pixel, a novel pixelation scheme—the line-scan method—is developed for this new detector. To implement line scanning, a multichannel high-voltage switching circuit and a multichannel charge-acquisition circuit are developed. These two circuits are controlled by an FPGA-based digital signal processing board, from which the information about the charge and position of each pixel can be sent to a PC. FPGA-based baseline compensation and switching noise rejection algorithms are used to improve the signal-to-noise ratio (SNR). The characteristic curve of the entire PXD system is acquired, and the correlation coefficients between the X-ray dose, and the signal intensity and the SNR were determined to be approximately 0.99 and 52.9, respectively. [ABSTRACT FROM AUTHOR]