Overestimation of free light chain antigen excess rate.

Authors :
Vercammen, Martine J.
Broodtaerts, Linda
Meirlaen, Patricia
Bossuyt, Xavier
Source :
Clinica Chimica Acta. Apr2015, Vol. 444, p297-302. 6p.
Publication Year :
2015

Abstract

Background Free light chains (FLC) are useful biomarkers for diagnosis and follow-up of plasma cell disorders. FLC quantification is encumbered by non-linearity and antigen excess (> 4-fold difference between results obtained at the 2000- and 100-fold dilution). Methods FLC concentration was measured with Freelite® reagents on the BNII, using 100- and 2000-fold dilutions in 3645 samples. Samples displaying antigen excess were re-measured at the 2000- and/or 400-fold dilution. Carryover was evaluated by tracing samples to cuvettes and by measuring a normal sample in cuvettes that previously contained samples with a high FLC concentration. Results Antigen excess occurred in 0.93% of samples for κ and in 0.55% of samples for λ . In 81.5% of the cases, it could not be confirmed by a re-analysis of a 2000-fold and/or a 400-fold diluted sample. Real antigen excess was documented in 0.25% and 0.03% of the samples for κ and λ FLC, respectively. In the low concentration range (2000-fold dilution), imprecision was high. False antigen excess was reduced by batch analysis, introducing cleaning and rinsing procedures and using the 400-fold dilution. No antigen excess was detected in samples with normal FLC concentrations. Conclusion Falsely high results occur by imprecision in the low concentration range and/or by carryover in cuvettes. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00098981
Volume :
444
Database :
Academic Search Index
Journal :
Clinica Chimica Acta
Publication Type :
Academic Journal
Accession number :
101917660
Full Text :
https://doi.org/10.1016/j.cca.2015.02.047