Back to Search Start Over

Design and Characterization of a Peltier-Cold Calibration Target for a 110-GHz Radiometer.

Authors :
Fernandez, Susana
Murk, Axel
Kampfer, Niklaus
Source :
IEEE Transactions on Geoscience & Remote Sensing. Jan2015, Vol. 53 Issue 1, p344-351. 8p.
Publication Year :
2015

Abstract

Blackbody targets are needed for the calibration of radiometers for remote sensing applications. The use of hot and cold references is essential for an accurate calibration. Commonly, the cold loads are made with liquid nitrogen, but for a campaign instrument, this is not a sustainable option. In this paper, we present the design, construction, and characterization of a new cold calibration target for a microwave radiometer operated at 110 GHz for the detection of pressure-broadened spectra from atmospheric ozone. The target is cooled with Peltier elements. Heat transfer simulations were made to optimize the design and minimize temperature gradients. Several tests were performed with a vector network analyzer, a radiometer, and a network of thermometers in order to characterize the blackbody built. Results show a homogeneous temperature along the absorber surface, notably low reflections, and a very high emissivity. [ABSTRACT FROM PUBLISHER]

Details

Language :
English
ISSN :
01962892
Volume :
53
Issue :
1
Database :
Academic Search Index
Journal :
IEEE Transactions on Geoscience & Remote Sensing
Publication Type :
Academic Journal
Accession number :
101187148
Full Text :
https://doi.org/10.1109/TGRS.2014.2322336