Back to Search Start Over

Current Monitoring in a Microchannel with Repeated Constrictions for Accurate Detection of Sample Location in Isotachophoresis.

Authors :
Karsenty, Merav
Rosenfeld, Tally
Gommed, Khaled
Bercovici, Moran
Source :
Analytical Chemistry. 1/6/2015, Vol. 87 Issue 1, p388-393. 3p.
Publication Year :
2015

Abstract

We present a new method for accurate detection of sample location in peak mode isotachophoresis (ITP). The technique is based on the design of a microchannel with multiple constrictions and on detecting the passage of the ITP interface through these constrictions. We achieve this by monitoring the electric current across the channel, which exhibits sharp decreases as the ITP interface moves more rapidly through the higher current density constrictions. We show that cross-correlation between the electric current signal and a predefined step function is an effective method for detecting changes in the slope of the electric current curve in real-time and is robust to changes in the composition of the buffers. We demonstrate the use of the technique to deliver sample to a designated location in the channel with an accuracy as low as 50 µm. Importantly, the method does not require the use of any optics and thus can be used to monitor the location of unlabeled species for application in a variety of ITP assays. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00032700
Volume :
87
Issue :
1
Database :
Academic Search Index
Journal :
Analytical Chemistry
Publication Type :
Academic Journal
Accession number :
101080132
Full Text :
https://doi.org/10.1021/ac5036346