Back to Search Start Over

On-site monitoring of uranium in low level liquid waste streams using U-Br-PADAP strip indicator paper technique.

Authors :
Ganesh, S.
Velavendan, P.
Pandey, N.
Kamachi Mudali, U.
Natarajan, R.
Source :
Journal of Radioanalytical & Nuclear Chemistry. Dec2014, Vol. 302 Issue 3, p1513-1518. 6p.
Publication Year :
2014

Abstract

A simple, rapid and selective colorimetric technique for the determination of low level concentration of uranium in nuclear waste streams is described. The proposed method is based on the intensity of U-Br-PADAP (Uranium-2-(5-Bromo-2-pyridylazo)-5-diethyl-aminophenol) colour formed on strip indicator paper. The effect of pH, concentration of chromogenic reagent, stability of the colour, volume of buffer solution and interference of metal ions were studied. The detection limit is 1 mg/L. Various uranium samples have been tested for the validation of the proposed method. The results obtained are reproducible and good agreement with those obtained by the standard technique. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02365731
Volume :
302
Issue :
3
Database :
Academic Search Index
Journal :
Journal of Radioanalytical & Nuclear Chemistry
Publication Type :
Academic Journal
Accession number :
100302894
Full Text :
https://doi.org/10.1007/s10967-014-3664-2