Back to Search Start Over

A Monolithic Segmented Germanium Detector with Highly Integrated Readout.

Authors :
Rumaiz, Abdul K.
Krings, T.
Siddons, D. P.
Kuczewski, A. J.
Protic, D.
Ross, C.
De Geronimo, G.
Zhong, Z.
Source :
IEEE Transactions on Nuclear Science. Dec2014 Part 2, Vol. 61 Issue 6, p3721-3726. 6p.
Publication Year :
2014

Abstract

We have constructed a pixelated germanium detector using a technique which has been shown to provide good isolation between adjacent pixels. In this work we present initial tests of the application of a low-noise CMOS ASIC to read out this detector. The detector has 64 pixels, each 0.5~\mm\times 5~\mm, arranged as a series of strips. It is connected by wire-bonds to two 32-channel application-specific integrated circuits (ASICs) which provide a complete photon-counting chain for every channel. Since the size of the pixel array is no longer restricted by the difficulties of instrumenting large channel-count conventional electronics, this development will open up the possibility of even larger arrays, similar to those offered by silicon detectors. [ABSTRACT FROM PUBLISHER]

Details

Language :
English
ISSN :
00189499
Volume :
61
Issue :
6
Database :
Academic Search Index
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
100077240
Full Text :
https://doi.org/10.1109/TNS.2014.2365358