1. Biological and Corrosion Evaluation of In Situ Alloyed NiTi Fabricated through Laser Powder Bed Fusion (LPBF)
- Author
-
Agnieszka Chmielewska, Anna Dobkowska, Ewa Kijeńska-Gawrońska, Michał Jakubczak, Agnieszka Krawczyńska, Emilia Choińska, Agnieszka Jastrzębska, David Dean, Bartłomiej Wysocki, and Wojciech Święszkowski
- Subjects
nitinol (nickel titanium) ,NiTi ,laser powder bed fusion (LBPF) ,corrosion ,ion release ,cytotoxicity ,Biology (General) ,QH301-705.5 ,Chemistry ,QD1-999 - Abstract
In this work, NiTi alloy parts were fabricated using laser powder bed fusion (LBPF) from pre-alloyed NiTi powder and in situ alloyed pure Ni and Ti powders. Comparative research on the corrosive and biological properties of both studied materials was performed. Electrochemical corrosion tests were carried out in phosphate buffered saline at 37 °C, and the degradation rate of the materials was described based on Ni ion release measurements. Cytotoxicity, bacterial growth, and adhesion to the surface of the fabricated coupons were evaluated using L929 cells and spherical Escherichia coli (E. coli) bacteria, respectively. The in situ alloyed NiTi parts exhibit slightly lower corrosion resistance in phosphate buffered saline solution than pre-alloyed NiTi. Moreover, the passive layer formed on in situ alloyed NiTi is weaker than the one formed on the NiTi fabricated from pre-alloyed NiTi powder. Furthermore, in situ alloyed NiTi and NiTi made from pre-alloyed powders have comparable cytotoxicity and biological properties. Overall, the research has shown that nitinol sintered using in situ alloyed pure Ni and Ti is potentially useful for biomedical applications.
- Published
- 2021
- Full Text
- View/download PDF