1. Development of new CdZnTe detectors for room-temperature high-flux radiation measurements.
- Author
-
Abbene, Leonardo, Gerardi, Gaetano, Raso, Giuseppe, Principato, Fabio, Zambelli, Nicola, Benassi, Giacomo, Bettelli, Manuele, and Zappettini, Andrea
- Subjects
- *
CADMIUM zinc telluride detectors , *X-ray spectroscopy technique , *X-ray imaging , *FLUX (Energy) , *ELECTROCHEMICAL electrodes - Abstract
Recently, CdZnTe (CZT) detectors have been widely proposed and developed for room-temperature X-ray spectroscopy even at high fluxes, and great efforts have been made on both the device and the crystal growth technologies. In this work, the performance of new travelling-heater-method (THM)-grown CZT detectors, recently developed at IMEM-CNR Parma, Italy, is presented. Thick planar detectors (3 mm thick) with gold electroless contacts were realised, with a planar cathode covering the detector surface (4.1 mm × 4.1 mm) and a central anode (2 mm × 2 mm) surrounded by a guard-ring electrode. The detectors, characterized by low leakage currents at room temperature (4.7 nA cm−2 at 1000 V cm−1), allow good room-temperature operation even at high bias voltages (>7000 V cm−1). At low rates (200 counts s−1), the detectors exhibit an energy resolution around 4% FWHM at 59.5 keV (241Am source) up to 2200 V, by using commercial front-end electronics (A250F/NF charge-sensitive preamplifier, Amptek, USA; nominal equivalent noise charge of 100 electrons RMS). At high rates (1 Mcounts s−1), the detectors, coupled to a custom-designed digital pulse processing electronics developed at DiFC of University of Palermo (Italy), show low spectroscopic degradations: energy resolution values of 8% and 9.7% FWHM at 59.5 keV (241Am source) were measured, with throughputs of 0.4% and 60% at 1 Mcounts s−1, respectively. An energy resolution of 7.7% FWHM at 122.1 keV (57Co source) with a throughput of 50% was obtained at 550 kcounts s−1 (energy resolution of 3.2% at low rate). These activities are in the framework of an Italian research project on the development of energy-resolved photon-counting systems for high-flux energy-resolved X-ray imaging. [ABSTRACT FROM AUTHOR]
- Published
- 2017
- Full Text
- View/download PDF