1. Evaluating intra‐fractional tumor motion in lung stereotactic radiotherapy with deep inspiration breath‐hold.
- Author
-
Fu, Weihua, Zhang, Yongqian, Mehta, Kiran, Chen, Alex, Musunuru, Hima Bindu, Pucci, Pietro, Kubis, Jason, and Huq, M. Saiful
- Subjects
STEREOTACTIC radiotherapy ,CONE beam computed tomography ,LUNG tumors ,LUNGS - Abstract
Purpose: To evaluate the intra‐fractional tumor motion in lung stereotactic body radiotherapy (SBRT) with deep inspiration breath‐hold (DIBH), and to investigate the adequacy of the current planning target volume (PTV) margins. Methods: Twenty‐eight lung SBRT patients with DIBH were selected in this study. Among the lesions, twenty‐three were at right or left lower lobe, two at right middle lobe, and three at right or left upper lobe. Post‐treatment gated cone‐beam computed tomography (CBCT) was acquired to quantify the intra‐fractional tumor shift at each treatment. These obtained shifts were then used to calculate the required PTV margin, which was compared with the current applied margin of 5 mm margin in anterior‐posterior (AP) and right‐left (RL) directions and 8 mm in superior‐inferior (SI) direction. The beam delivery time was prolonged with DIBH. The actual beam delivery time with DIBH (Tbeam_DIBH) was compared with the beam delivery time without DIBH (Tbeam_wo_DIBH) for the corresponding SBRT plan. Results: A total of 113 treatments were analyzed. At six treatments (5.3%), the shifts exceeded the tolerance defined by the current PTV margin. The average shifts were 0.0 ± 1.9 mm, 0.1±1.5 mm, and ‐0.5 ± 3.7 mm in AP, RL, and SI directions, respectively. The required PTV margins were determined to be 4.5, 3.9, and 7.4 mm in AP, RL, and SI directions, respectively. The average Tbeam_wo_DIBH and Tbeam_DIBH were 2.4 ± 0.4 min and 3.6 ± 1.5 min, respectively. The average treatment slot for lung SBRT with DIBH was 25.3 ± 7.9 min. Conclusion: Intra‐fractional tumor motion is the predominant source of treatment uncertainties in CBCT‐guided lung SBRT with DIBH. The required PTV margin should be determined based on data specific to each institute, considering different techniques and populations. Our data indicate that our current applied PTV margin is adequate, and it is possible to reduce further in the RL direction. The time increase of Tbeam_DIBH, relative to the treatment slot, is not clinically significant. [ABSTRACT FROM AUTHOR]
- Published
- 2024
- Full Text
- View/download PDF