1. Randomized crossover trial of a demand oxygen delivery system in nocturnal hypoxemia
- Author
-
Atsuhiko Yatani, Tatsuya Nagano, Sae Murakami, Takehiro Otoshi, Daisuke Hazama, Naoko Katsurada, Masatsugu Yamamoto, Motoko Tachihara, Yoshihiro Nishimura, and Kazuyuki Kobayashi
- Subjects
Chronic respiratory failure ,Home oxygen therapy ,Auto-demand device ,Stationary device ,Medicine ,Science - Abstract
Abstract The newly developed portable oxygen concentrator with an auto-demand oxygen delivery system (auto-DODS) automatically switches between 3 sensitivities according to the negative pressure gradient of inhalation and supplies oxygen only during inhalation. The aim of this study was to verify the effectiveness and safety of auto-demand devices compared with a continuous flow oxygen concentrator, during sleep, in a randomized crossover noninferiority trial. We alternatively used an auto-DODS or a continuous-flow oxygen concentrator, all night on separate days for HOT (Home Oxygen Therapy) patients with nocturnal hypoxemic symptoms. The primary endpoints were the mean value of oxygen saturation (SpO2) over the total sleep time. The secondary endpoints included the efficacy endpoints and the safety endpoints. Regarding the primary endpoint, the difference in SpO2 between the auto-DODS and continuous flow was 0.835%. Since the upper limit of this difference did not exceed 2.8, which was set as the noninferiority margin, it was shown that the auto-DODS did not reduce SpO2 by at least 2.8% on average compared to continuous flow. No equipment failure or exacerbation of disease was observed, confirming the safety of the auto-DODS during the night.
- Published
- 2024
- Full Text
- View/download PDF