1. Accurate determination of high sulfur content in sulfide samples: an optimized ICP-OES method
- Author
-
Lin Zhang, Min Yang, Dong Wang, Kaixin Yu, Zhuoyun Chen, Xue Bing, Peng Zhao, Yiqing Bao, and Xuebo Yin
- Subjects
Sulfide ,Sulfur content ,ICP-OES ,Anti-corrosion acid digestion bomb ,Chemistry ,QD1-999 ,Analytical chemistry ,QD71-142 - Abstract
Abstract Determination of high sulfur content in sulfide minerals is crucial for various applications, including geochemistry, mining, and environmental monitoring. While inductively coupled plasma optical emission spectrometry (ICP-OES) is the preferred method for sulfur analysis, conventional sample preparation techniques often fall short when dealing with high sulfur concentrations. This study investigated optimized digestion conditions for accurate sulfur determination in sulfide minerals using ICP-OES. We employed anti-corrosion acid digestion bombs, an aqua regia–hydrofluoric acid system, and a digestion temperature of 180 °C, achieving exceptional results. Certified reference material recoveries ranged from 99.9 to 100.5%, with relative standard deviations between 0.42 and 0.84%. These results demonstrate the high accuracy and precision of the method. The optimized digestion protocol presented in this study provides a robust and reliable approach for accurate sulfur analysis in sulfide minerals, particularly for high concentrations.
- Published
- 2024
- Full Text
- View/download PDF