1. Simplified method for creating a density-absorbed dose calibration curve for the low dose range from Gafchromic EBT3 film
- Author
-
Tatsuhiro Gotanda, Toshizo Katsuda, Rumi Gotanda, Tadao Kuwano, Takuya Akagawa, Nobuyoshi Tanki, Akihiko Tabuchi, Tetsunori Shimono, and Yasuyuki Kawaji
- Subjects
Calibration curve ,diagnostic dose range ,radiochromic film ,step-shaped Al filter ,Medical physics. Medical radiology. Nuclear medicine ,R895-920 - Abstract
Radiochromic film dosimeters have a disadvantage in comparison with an ionization chamber in that the dosimetry process is time-consuming for creating a density-absorbed dose calibration curve. The purpose of this study was the development of a simplified method of creating a density-absorbed dose calibration curve from radiochromic film within a short time. This simplified method was performed using Gafchromic EBT3 film with a low energy dependence and step-shaped Al filter. The simplified method was compared with the standard method. The density-absorbed dose calibration curves created using the simplified and standard methods exhibited approximately similar straight lines, and the gradients of the density-absorbed dose calibration curves were −32.336 and −33.746, respectively. The simplified method can obtain calibration curves within a much shorter time compared to the standard method. It is considered that the simplified method for EBT3 film offers a more time-efficient means of determining the density-absorbed dose calibration curve within a low absorbed dose range such as the diagnostic range.
- Published
- 2016
- Full Text
- View/download PDF