1. Comparisons between In-Check DIAL® and PF810® in evaluation and training inspiratory capacity for using dry powder inhalers
- Author
-
Yixing Wu, Li Li, Ying Gong, Xingjing Li, Xiaofen Ye, and Jing Zhang
- Subjects
Chronic obstructive pulmonary diseases ,Asthma ,Inhalers ,Inspiratory capacity ,Peak inspiratory flow ,Medical devices ,Diseases of the respiratory system ,RC705-779 - Abstract
Abstract Background Dry powder inhalers (DPIs) rely on both internal resistance and patients’ inspiratory capacity for effective operation. Optimal inspiratory technique is crucial for DPI users. This study assessed the accuracy and repeatability of two available devices, PF810® and In-Check DIAL®, and analyzed their measurement errors and consistency in detecting inspiratory capacity. Methods The accuracy and repeatability of peak inspiratory flow (PIF) and forced inspiratory vital capacity (FIVC) against various internal resistances of the two devices were assessed using standard waveforms generated by a breathing simulator. The agreement of PIF measurements between the two devices in healthy volunteers and chronic obstructive pulmonary disease (COPD) patients was analyzed with the intraclass correlation coefficient and Bland–Altman graphical analysis. Results PF810® showed great accuracy and repeatability in measuring PIF, except for square waveforms at the lowest flow rate (20 L/min). In-Check DIAL® exhibited poor accuracy against high resistance levels. In scenarios with no resistance, In-Check DIAL® had significantly smaller measurement errors than PF810®, but larger errors against high resistance levels. The two devices showed excellent agreement (ICC > 0.80, P
- Published
- 2024
- Full Text