1. An augmented hybrid multibaseline and referenceless MR thermometry motion compensation algorithm for MRgHIFU hyperthermia.
- Author
-
Wong, Suzanne M., Akbulatov, Arthur, Macsemchuk, Craig A., Headrick, Andrew, Phoebe Luo, Drake, James M., and Waspe, Adam C.
- Subjects
THERMOMETRY ,PERIODIC motion ,IMAGING phantoms ,FEVER ,PRINCIPAL components analysis - Abstract
Purpose: A hybrid principal component analysis and projection onto dipole fields (PCA-PDF) MR thermometry motion compensation algorithm was optimized with atlas image augmentation and validated. Methods: Experiments were conducted on a 3T Philips MRI and Profound V1 Sonalleve high intensity focused ultrasound (high intensity focused ultrasound system. An MR-compatible robot was configured to induce motion on custom gelatin phantoms. Trials with periodic and sporadic motion were introduced on phantoms while hyperthermia was administered. The PCA-PDF algorithm was augmented with a predictive atlas to better compensate for larger sporadic motion. Results: During periodic motion, the temperature SD in the thermometry was improved from 1.1 ± 0.1 to 0.5 ± 0.1°C with both the original and augmented PCA-PDF application. For large sporadic motion, the augmented atlas improved the motion compensation from the original PCA-PDF correction from 8.8 ± 0.5 to 0.7 ± 0.1°C. Conclusion: The PCA-PDF algorithm improved temperature accuracy to <1°C during periodic motion, butwas not able to adequately address sporadicmotion. By augmenting the PCA-PDF algorithm, temperature SD during large sporadic motion was also reduced to <1°C, greatly improving the original PCA-PDF algorithm. [ABSTRACT FROM AUTHOR]
- Published
- 2024
- Full Text
- View/download PDF