1. Clinical and microbiological effect of pulsed xenon ultraviolet disinfection to reduce multidrug-resistant organisms in the intensive care unit in a Japanese hospital: a before-after study
- Author
-
Keita Morikane, Shoko Suzuki, Jun Yoshioka, Jun Yakuwa, Masaki Nakane, and Kenji Nemoto
- Subjects
Ultraviolet disinfection device ,Multidrug-resistant organism ,Acinetobacter baumannii ,Environmental disinfection ,Infectious and parasitic diseases ,RC109-216 - Abstract
Abstract Background No-touch environmental disinfection using ultraviolet devices has been highlighted in the past several years to control the transmission of multidrug-resistant organisms (MDROs). However, its effectiveness in non-US healthcare settings is yet to be examined. This study aimed to evaluate the effectiveness of disinfection by portable pulsed xenon ultraviolet (PX-UV) devices in controlling transmission of MDROs in a non-US healthcare setting. Methods All patients admitted in the intensive care unit in a 629-bed tertiary referral hospital in Japan from August 2016 to February 2019 were enrolled. During the study period, PX-UV disinfection was added to manual terminal cleaning after every patient transfer/discharge. For microbiological evaluation, surfaces were selected for sampling by contact plates before/after manual cleaning and after PX-UV. After overnight incubation, colonies on the plates were counted. Results The incidence of newly acquired methicillin-resistant Staphylococcus aureus (MRSA) declined significantly (13.8 to 9.9 per 10,000 patient days, incidence rate ratio 0.71, p = 0.002), as well as that of newly acquired drug-resistant Acinetobacter (48.5 to 18.1, 0.37, p