1. Evaluation of the Cue Health point-of-care COVID-19 (SARS-CoV-2 nucleic acid amplification) test at a community drive through collection center.
- Author
-
Donato LJ, Trivedi VA, Stransky AM, Misra A, Pritt BS, Binnicker MJ, and Karon BS
- Subjects
- Carrier State, Humans, Minnesota, Point-of-Care Systems, Prospective Studies, Sensitivity and Specificity, Specimen Handling instrumentation, COVID-19 Nucleic Acid Testing methods, Nasopharynx virology, Nucleic Acid Amplification Techniques methods, Specimen Handling methods
- Abstract
Point-of-care (POC) tests are in high demand in order to facilitate rapid care decisions for patients suspected of SARS-CoV-2. We conducted a clinical validation study of the Cue Health POC nucleic acid amplification test (NAAT) using the Cue lower nasal swab, compared to a reference NAAT using standard nasopharyngeal swab, in 292 symptomatic and asymptomatic outpatients for SARS-CoV-2 detection in a community drive through collection setting. Positive percent agreement between Cue COVID-19 and reference SARS-CoV-2 test was 91.7% (22 of 24); or 95.7% (22 of 23) when one patient with no tie-breaker method was excluded. Negative percent agreement was 98.4% (239 of 243), and there were 25 (8.6%) invalid or canceled results. The Cue COVID-19 test demonstrated very good positive and negative percent agreement with central laboratory tests and will be useful in settings where accurate POC testing is needed to facilitate management of patients suspected of COVID-19., (Copyright © 2021 Elsevier Inc. All rights reserved.)
- Published
- 2021
- Full Text
- View/download PDF