1. Racial bias and reproducibility in pulse oximetry among medical and surgical inpatients in general care in the Veterans Health Administration 2013-19: multicenter, retrospective cohort study
- Author
-
Valeria S M Valbuena, Sarah Seelye, Michael W Sjoding, Thomas S Valley, Robert P Dickson, Steven E Gay, Dru Claar, Hallie C Prescott, and Theodore J Iwashyna
- Subjects
Adult ,Inpatients ,Racism ,Humans ,Reproducibility of Results ,Veterans Health ,General Medicine ,Oximetry ,Hypoxia ,Retrospective Studies - Abstract
ObjectivesTo evaluate measurement discrepancies by race between pulse oximetry and arterial oxygen saturation (as measured in arterial blood gas) among inpatients not in intensive care.DesignMulticenter, retrospective cohort study using electronic medical records from general care medical and surgical inpatients.SettingVeteran Health Administration, a national and racially diverse integrated health system in the United States, from 2013 to 2019.ParticipantsAdult inpatients in general care (medical and surgical), in Veteran Health Administration medical centers.Main outcomes measuresOccult hypoxemia (defined as arterial blood oxygen saturation (SaO2) of 2) reading of ≥92%), and whether rates of occult hypoxemia varied by race and ethnic origin.ResultsA total of 30 039 pairs of SpO2-SaO2readings made within 10 minutes of each other were identified during the study. These pairs were predominantly among non-Hispanic white (21 918 (73.0%)) patients; non-Hispanic black patients and Hispanic or Latino patients accounted for 6498 (21.6%) and 1623 (5.4%) pairs in the sample, respectively. Among SpO2values greater or equal to 92%, unadjusted probabilities of occult hypoxemia were 15.6% (95% confidence interval 15.0% to 16.1%) in white patients, 19.6% (18.6% to 20.6%) in black patients (Pvwhite patients, with similar P values in adjusted models), and 16.2% (14.4% to 18.1%) in Hispanic or Latino patients (P=0.53vwhite patients, P2-SaO2pairs restricted to occur within 5 minutes and 2 minutes. In white patients, an initial SpO2-SaO2pair with little difference in saturation was associated with a 2.7% (95% confidence interval −0.1% to 5.5%) probability of SaO22-SaO2reading showing an SpO2of 92%, but black patients had a higher probability (12.9% (−3.3% to 29.0%)).ConclusionsIn general care inpatient settings across the Veterans Health Administration where paired readings of arterial blood gas (SaO2) and pulse oximetry (SpO2) were obtained, black patients had higher odds than white patients of having occult hypoxemia noted on arterial blood gas but not detected by pulse oximetry. This difference could limit access to supplemental oxygen and other more intensive support and treatments for black patients.
- Published
- 2022