1. A directional total variation minimization algorithm for isotropic resolution in digital breast tomosynthesis
- Author
-
Sidky, Emil Y., Wu, Xiangyi, Duan, Xiaoyu, Huang, Hailing, Zhao, Wei, Zhang, Leo Y., Phillips, John Paul, Zhang, Zheng, Chen, Buxin, Xia, Dan, Reiser, Ingrid S., and Pan, Xiaochuan
- Subjects
Physics - Medical Physics - Abstract
An optimization-based image reconstruction algorithm is developed for contrast enhanced digital breast tomosynthesis (DBT) using dual-energy scanning. The algorithm minimizes directional total variation (TV) with a data discrepancy and non-negativity constraints. Iodinated contrast agent (ICA) imaging is performed by reconstructing images from dual-energy DBT data followed by weighted subtraction. Physical DBT data is acquired with a Siemens Mammomat scanner of a structured breast phantom with ICA inserts. Results are shown for both directional TV minimization and filtered back-projection for reference. It is seen that directional TV is able to substantially reduce depth blur for the ICA objects., Comment: Proceedings paper for accepted contribution to the 8th International Conference on Image Formation in X-Ray Computed Tomography (https://www.ct-meeting.org)
- Published
- 2024