1. DICOM CT-based Monte Carlo dose calculations for prostate permanent implant using 125I seed sources.
- Author
-
Mishra, Subhalaxmi, Deshpande, Sudesh, S. Pathan, Munir, and Selvam, T. Palani
- Subjects
- *
IMAGING phantoms , *RADIOISOTOPE brachytherapy , *COMPUTED tomography , *PROSTATE , *SEEDS - Abstract
Dose distributions are calculated for a clinical prostate implant consisting of multiple 125I sources using Monte Carlo-based egs_brachy code. A voxelised heterogeneous CT phantom, generated from DICOM CT images is used. Dose distributions vary significantly with scoring and transport media. About 3% interseed attenuation and 12% dose differences between prostate tissue and water are observed at D 90. Monte Carlo-calculated dose values are smaller as compared to the TG-43 formalism-based TPS-generated dose values at a given location. Hence, patient-specific model-based dose calculation algorithm is preferred over TG-43 formalism-based TPS. • Dose distributions are studied for a prostate implant consisting of 57 numbers of indigenous 125I brachytherapy sources. • Voxelised heterogeneous CT phantom, generated from DICOM CT images is used. • Demonstrated the capability of a new Monte Carlo code egs_brachy. • Interseed attenuation and tissue heterogeneity are also quantified for indigenous 125I source. • It is found that Monte Carlo-calculated dose distributions are smaller as compared to TPS calculated one. [ABSTRACT FROM AUTHOR]
- Published
- 2023
- Full Text
- View/download PDF