1. Comparison of leakage at the implant to abutment connection in several implants types using a gas flow method
- Author
-
JH Torres, M. Mechali, M.-A. Fauroux, O. Romieu, C. Biens, C Anxionnat, Laboratoire de Bioingénierie et NanoSciences (LBN), and Université Montpellier 1 (UM1)-Université de Montpellier (UM)
- Subjects
03 medical and health sciences ,0302 clinical medicine ,Otorhinolaryngology ,0206 medical engineering ,Surgery ,[SDV.MHEP.CHI]Life Sciences [q-bio]/Human health and pathology/Surgery ,030206 dentistry ,02 engineering and technology ,General Medicine ,Oral Surgery ,020601 biomedical engineering ,ComputingMilieux_MISCELLANEOUS ,[SDV.MHEP]Life Sciences [q-bio]/Human health and pathology - Abstract
Resume Introduction L’objet de cette etude a ete de comparer la fuite de la connexion implantaire pour plusieurs types d’implants, par une nouvelle methode de diffusion gazeuse. Materiel et methodes Soixante-huit implants parmi 13 types differents ont ete compares. La fuite par diffusion d’azote a ete mesuree apres serrage des connexions aux couples preconises par les fabricants. Resultats Une difference significative d’hermeticite a ete notee entre les divers types d’implant. Cette difference n’est expliquee ni par la conformation des diverses connexions (plates, coniques) ni par les divers couples de serrage preconises par les constructeurs. Discussion Les auteurs suggerent que la difference d’hermeticite entre les divers systemes implantaires pourrait etre liee essentiellement a la qualite et a la precision de l’usinage.