1. High risk of elevated metal concentrations with 9/10-mm stem trunnions and highly cross-linked polyethylene grafted with poly(2-methacryloyloxyethyl phosphorylcholine) in total hip arthroplasty
- Author
-
Kyosuke Kobayashi, Kenichi Kidera, Kazuteru Shiraishi, Narihiro Okazaki, Ko Chiba, Akihiko Yonekura, and Makoto Osaki
- Subjects
Trunnion ,Fretting and corrosion ,Serum metal concentrations ,Adverse local tissue reaction ,Metal-on-polyethylene ,Orthopedic surgery ,RD701-811 ,Diseases of the musculoskeletal system ,RC925-935 - Abstract
Abstract Background The risks of metal release due to fretting and corrosion at the head–neck junction and consequent adverse local tissue reaction (ALTR) have concerns in metal-on-polyethylene (MoP) total hip arthroplasty (THA). Although trunnions have become thinner in diameter to increase the range of motion, it has remained unclear whether this change affects metal release and ALTR in vivo. This study aimed to investigate serum metal concentrations and the prevalence of ALTR in MoP THA with a 9/10-mm stem trunnion. Patients and methods A consecutive series of 37 hips that underwent THA using MoP grafted with poly(2-methacryloyloxyethyl phosphorylcholine) (PMPC) with a 9/10-mm trunnion stem were retrospectively reviewed. Serum metal levels were assessed and compared with those in MoP THA with a 10/12-mm trunnion stem. ALTR was diagnosed with serum metal levels and cross-sectional images. The factors associated with serum metal levels were also assessed. Results The median serum cobalt and chromium levels were 1.5 μg/L and 1.0 μg/L in the 9/10-mm group and 0.2 μg/L and 0.4 μg/L in the 10/12-mm group, respectively. ALTR was found in 5 hips of 3 patients. Revision surgery was performed in 4 hips, and all stem trunnions and femoral heads showed severe corrosion. Postoperative walking ability was associated with serum metal levels. Conclusion It was found that a 9/10-mm stem trunnion with MoP grafted with PMPC had high risks of metal release in primary THA. Careful follow-up and cross-sectional imaging are needed to detect ALTR for early revision.
- Published
- 2023
- Full Text
- View/download PDF