1. Comparability of Results Between a Point-of-Care and an Automated Instrument for Measurement of B-Type Natriuretic Peptide
- Author
-
Shah, Kevin, Terracciano, Garrett J, Jiang, Kevin, Maisel, Alan S, and Fitzgerald, Robert L
- Subjects
i-STAT ,Architect ,Heart Failure ,Natriuretic Peptide ,BNP ,Medicine ,Medical emergencies. Critical care. Intensive care. First aid ,RC86-88.9 - Abstract
Objectives: Heart failure is one of the leading causes of death in the U.S. The incorporation of B-type natriuretic peptide (BNP) measurements when triaging patients presenting with shortness of breath has improved the diagnostic and prognostic ability of physicians. Currently, there are no point-of-care systems for quantifying BNP that can be used without sacrificing accuracy. We compared the analytical performance of the Abbott i-STAT analyzer, a handheld point-of-care system for measuring BNP, with the lab-based system, the Abbott ARCHITECT.Methods: One-hundred fifty samples were collected from three clinical settings: 41 from the Emergency Department, 58 from the inpatient wards, and 51 from heart failure outpatient clinics. Linear regression and bias difference analyses were run to evaluate the accuracy of the i-STAT. Correlation between the i-STAT and Architect BNP values were made with values of BNP.Results: The correlation coefficient was r=0.977 (N=150, p
- Published
- 2010