10 results on '"Knight, R. T."'
Search Results
2. In celebration of my Patrick days.
- Author
-
Knight, R. T.
- Subjects
- *
RETINOSCOPY , *EYE examination - Abstract
The article discusses the author's experience of using a Patrick retinoscope in his refractive examination practice.
- Published
- 2007
3. A practical and theoretical definition of very small field size for radiotherapy output factor measurements.
- Author
-
Charles, P. H., Cranmer‐Sargison, G., Thwaites, D. I., Crowe, S. B., Kairn, T., Knight, R. T., Kenny, J., Langton, C. M., and Trapp, J. V.
- Subjects
- *
RADIOTHERAPY , *DETECTORS , *MONTE Carlo method , *UNCERTAINTY , *ECONOMIC equilibrium , *RADIATION dosimetry - Abstract
Purpose: This work introduces the concept of very small field size. Output factor (OPF) measurements at these field sizes require extremely careful experimental methodology including the measurement of dosimetric field size at the same time as each OPF measurement. Two quantifiable scientific definitions of the threshold of very small field size are presented. Methods: A practical definition was established by quantifying the effect that a 1 mm error in field size or detector position had on OPFs and setting acceptable uncertainties on OPF at 1%. Alternatively, for a theoretical definition of very small field size, the OPFs were separated into additional factors to investigate the specific effects of lateral electronic disequilibrium, photon scatter in the phantom, and source occlusion. The dominant effect was established and formed the basis of a theoretical definition of very small fields. Each factor was obtained using Monte Carlo simulations of a Varian iX linear accelerator for various square field sizes of side length from 4 to 100 mm, using a nominal photon energy of 6 MV. Results: According to the practical definition established in this project, field sizes =15 mm were considered to be very small for 6 MV beams for maximal field size uncertainties of 1 mm. If the acceptable uncertainty in the OPF was increased from 1.0% to 2.0%, or field size uncertainties are 0.5 mm, field sizes =12 mm were considered to be very small. Lateral electronic disequilibrium in the phantom was the dominant cause of change in OPF at very small field sizes. Thus the theoretical definition of very small field size coincided to the field size at which lateral electronic disequilibrium clearly caused a greater change in OPF than any other effects. This was found to occur at field sizes =12 mm. Source occlusion also caused a large change in OPF for field sizes =8 mm. Based on the results of this study, field sizes =12 mm were considered to be theoretically very small for 6 MV beams. Conclusions: Extremely careful experimental methodology including the measurement of dosimetric field size at the same time as output factor measurement for each field size setting and also very precise detector alignment is required at field sizes at least =12 mm and more conservatively =15 mm for 6 MV beams. These recommendations should be applied in addition to all the usual considerations for small field dosimetry, including careful detector selection. [ABSTRACT FROM AUTHOR]
- Published
- 2014
- Full Text
- View/download PDF
4. Monte Carlo-based diode design for correction-less small field dosimetry.
- Author
-
Charles, P. H., Crowe, S. B., Kairn, T., Knight, R. T., Hill, B., Kenny, J., Langton, C. M., and Trapp, J. V.
- Subjects
- *
DIODES , *RADIATION dosimetry , *MONTE Carlo method , *PHOTONS , *SILICON - Abstract
Due to their small collecting volume, diodes are commonly used in small field dosimetry. However, the relative sensitivity of a diode increases with decreasing small field size. Conversely, small air gaps have been shown to cause a significant decrease in the sensitivity of a detector as the field size is decreased. Therefore, this study uses Monte Carlo simulations to look at introducing air upstream to diodes such that they measure with a constant sensitivity across all field sizes in small field dosimetry. Varying thicknesses of air were introduced onto the upstream end of two commercial diodes (PTW 60016 photon diode and PTW 60017 electron diode), as well as a theoretical unenclosed silicon chip using field sizes as small as 5 mm × 5 mm. The metric Dw,Q/DDet,Q used in this study represents the ratio of the dose to a point of water to the dose to the diode active volume, for a particular field size and location. The optimal thickness of air required to provide a constant sensitivity across all small field sizes was found by plotting Dw,Q/DDet,Q as a function of introduced air gap size for various field sizes, and finding the intersection point of these plots. That is, the point at which Dw,Q/DDet,Q was constant for all field sizes was found. The optimal thickness of air was calculated to be 3.3, 1.15 and 0.10 mm for the photon diode, electron diode and unenclosed silicon chip, respectively. The variation in these results was due to the different design of each detector. When calculated with the new diode design incorporating the upstream air gap, k fclin, fmsr Qclin,Qmsr was equal to unity to within statistical uncertainty (0.5%) for all three diodes. Cross-axis profilemeasurements were also improved with the new detector design. The upstream air gap could be implanted on the commercial diodes via a cap consisting of the air cavity surrounded by water equivalent material. The results for the unclosed silicon chip show that an ideal small field dosimetry diode could be created by using a silicon chip with a small amount of air above it [ABSTRACT FROM AUTHOR]
- Published
- 2013
- Full Text
- View/download PDF
5. Executive functions after orbital or lateral prefrontal lesions: Neuropsychological profiles and self-reported executive functions in everyday living.
- Author
-
Løvstad, M., Funderud, I., Endestad, T., Due-Tønnessen, P., Meling, T. R., Lindgren, M., Knight, R. T., and Solbakk, A. K.
- Subjects
- *
ANALYSIS of covariance , *ANALYSIS of variance , *BRAIN injuries , *COGNITION , *STATISTICAL correlation , *PSYCHOLOGICAL tests , *RESEARCH funding , *STATISTICS , *DATA analysis , *CASE-control method , *DATA analysis software - Abstract
Objective: This study examined the effects of chronic focal lesions to the lateral prefrontal cortex (LPFC) or orbitofrontal cortex (OFC) on neuropsychological test performance and self-reported executive functioning in everyday living. Methods: Fourteen adults with OFC lesions were compared to 10 patients with LPFC injuries and 21 healthy controls. Neuropsychological tests with emphasis on measures of cognitive executive function were administered along with the Behavior Rating Inventory of Executive Functions (BRIEF-A) and a psychiatric screening instrument. Results: The LPFC group differed from healthy controls on neuropsychological tests of sustained mental effort, response inhibition, working memory and mental switching, while the BRIEF-A provided more clinically important information on deficits in everyday life in the OFC group compared to the LPFC group. Correlations between neuropsychological test results and BRIEF-A were weak, while the BRIEF-A correlated strongly with emotional distress. Conclusions: It was demonstrated that LPFC damage is particularly prone to cause cognitive executive deficit, while OFC injury is more strongly associated with self-reported dysexecutive symptoms in everyday living. The study illustrates the challenge of identifying executive deficit in individual patients and the lack of strong anatomical specificity of the currently employed methods. There is a need for an integrative methodological approach where standard testing batteries are supplemented with neuropsychiatric and frontal-specific rating scales. [ABSTRACT FROM AUTHOR]
- Published
- 2012
- Full Text
- View/download PDF
6. The effect of very small air gaps on small field dosimetry.
- Author
-
Charles, P. H., Crowe, S. B., Kairn, T., Kenny, J., Lehmann, J., Lye, J., Dunn, L., Hill, B., Knight, R. T., Langton, C. M., and Trapp, J. V.
- Subjects
- *
RADIATION dosimetry , *STEREOTAXIC techniques , *LINEAR accelerators , *MONTE Carlo method , *DETECTORS , *LUMINESCENCE , *IMAGING phantoms - Abstract
The purpose of this study was to investigate the effect of very small air gaps (less than 1 mm) on the dosimetry of small photon fields used for stereotactic treatments. Measurements were performed with optically stimulated luminescent dosimeters (OSLDs) for 6 MV photons on a Varian 21iX linear accelerator with a Brainlab µMLC attachment for square field sizes down to 6 mm × 6 mm. Monte Carlo simulations were performed using EGSnrc C++ user code cavity. It was found that the Monte Carlo model used in this study accurately simulated the OSLD measurements on the linear accelerator. For the 6 mm field size, the 0.5 mm air gap upstream to the active area of the OSLD caused a 5.3% dose reduction relative to a Monte Carlo simulation with no air gap. A hypothetical 0.2 mm air gap caused a dose reduction >2%, emphasizing the fact that even the tiniest air gaps can cause a large reduction in measured dose. The negligible effect on an 18 mm field size illustrated that the electronic disequilibrium caused by such small air gaps only affects the dosimetry of the very small fields. When performing small field dosimetry, care must be taken to avoid any air gaps, as can be often present when inserting detectors into solid phantoms. It is recommended that very small field dosimetry is performed in liquid water. When using small photon fields, sub-millimetre air gaps can also affect patient dosimetry if they cannot be spatially resolved on a CT scan. However the effect on the patient is debatable as the dose reduction caused by a 1 mm air gap, starting out at 19% in the first 0.1 mm behind the air gap, decreases to <5% after just 2 mm, and electronic equilibrium is fully re-established after just 5 mm. [ABSTRACT FROM AUTHOR]
- Published
- 2012
- Full Text
- View/download PDF
7. Monte Carlo verification of gel dosimetry measurements for stereotactic radiotherapy.
- Author
-
Kairn, T., Taylor, M. L., Crowe, S. B., Dunn, L., Franich, R. D., Kenny, J., Knight, R. T., and Trapp, J. V.
- Subjects
- *
MONTE Carlo method , *RADIATION dosimetry , *STEREOTAXIC techniques , *RADIOTHERAPY , *RADIOSURGERY , *SIMULATION methods & models , *COMPARATIVE studies - Abstract
The quality assurance of stereotactic radiotherapy and radiosurgery treatments requires the use of small-field dose measurements that can be experimentally challenging. This study used Monte Carlo simulations to establish that PAGAT dosimetry gel can be used to provide accurate, high-resolution, three-dimensional dose measurements of stereotactic radiotherapy fields. A small cylindrical container (4 cm height, 4.2 cm diameter) was filled with PAGAT gel, placed in the parietal region inside a CIRS head phantom and irradiated with a 12-field stereotactic radiotherapy plan. The resulting threedimensional dose measurement was read out using an optical CT scanner and compared with the treatment planning prediction of the dose delivered to the gel during the treatment. A BEAMnrc/DOSXYZnrc simulation of this treatment was completed, to provide a standard against which the accuracy of the gel measurement could be gauged. The three-dimensional dose distributions obtained from Monte Carlo and from the gel measurement were found to be in better agreement with each other than with the dose distribution provided by the treatment planning system's pencil beam calculation. Both sets of data showed close agreement with the treatment planning system's dose distribution through the centre of the irradiated volume and substantial disagreement with the treatment planning system at the penumbrae. The Monte Carlo calculations and gel measurements both indicated that the treated volume was up to 3 mm narrower, with steeper penumbrae and more variable out-of-field dose, than predicted by the treatment planning system. The Monte Carlo simulations allowed the accuracy of the PAGAT gel dosimeter to be verified in this case, allowing PAGAT gel to be utilized in themeasurement of dose from stereotactic and other radiotherapy treatments, with greater confidence in the future. [ABSTRACT FROM AUTHOR]
- Published
- 2012
- Full Text
- View/download PDF
8. Technical Note: Modeling a complex micro-multileaf collimator using the standard BEAMnrc distribution.
- Author
-
Kairn, T., Kenny, J., Crowe, S. B., Fielding, A. L., Franich, R. D., Johnston, P. N., Knight, R. T., Langton, C. M., Schlect, D., and Trapp, J. V.
- Subjects
- *
RADIOTHERAPY , *MONTE Carlo method , *COLLIMATORS , *IONIZATION chambers , *RADIOSURGERY - Abstract
Purpose: The component modules in the standard BEAMnrc distribution may appear to be insufficient to model micro-multileaf collimators that have trifaceted leaf ends and complex leaf profiles. This note indicates, however, that accurate Monte Carlo simulations of radiotherapy beams defined by a complex collimation device can be completed using BEAMnrc’s standard VARMLC component module. Methods: That this simple collimator model can produce spatially and dosimetrically accurate microcollimated fields is illustrated using comparisons with ion chamber and film measurements of the dose deposited by square and irregular fields incident on planar, homogeneous water phantoms. Results: Monte Carlo dose calculations for on-axis and off-axis fields are shown to produce good agreement with experimental values, even on close examination of the penumbrae. Conclusions: The use of a VARMLC model of the micro-multileaf collimator, along with a commissioned model of the associated linear accelerator, is therefore recommended as an alternative to the development or use of in-house or third-party component modules for simulating stereotactic radiotherapy and radiosurgery treatments. Simulation parameters for the VARMLC model are provided which should allow other researchers to adapt and use this model to study clinical stereotactic radiotherapy treatments. [ABSTRACT FROM AUTHOR]
- Published
- 2010
- Full Text
- View/download PDF
9. The Neural Substrates of Visual Implicit Memory: Do the Two Hemispheres Play Different Roles?
- Author
-
Kroll, N. E. A., Yonelinas, A. P., Kishiyama, M. M., Baynes, K., Knight, R. T., and Gazzaniga, M. S.
- Subjects
- *
MEMORY , *OCCIPITAL lobe , *CEREBRAL cortex , *RECOGNITION (Psychology) - Abstract
Identification of visually presented words is facilitated by implicit memory, or visual priming, for past visual experiences with those words. There is disagreement over the neuro-anatomical substrates of this form of implicit memory. Several studies have suggested that this form of priming relies on a visual word-form system localized in the right occipital lobe, whereas other studies have indicated that both hemispheres are equally involved. The discrepancies may be related to the types of priming tasks that have been used because the former studies have relied primarily on word-stem completion tasks and the latter on tasks like word-fragment completion. The present experiments compared word-fragment and word-stem measurements of visual implicit memory in patients with right occipital lobe lesions and patients with complete callosotomies. The patients showed normal visual implicit memory on fragment completion tests, but essentially no visual priming on standard stem completion tests. However, when we used a set of word stems that had only one correct solution for each test item, as was true of the items in the fragment completion tests, the patients showed normal priming effects. The results indicate that visual implicit memory for words is not solely dependent upon the right hemisphere, rather it reflects changes in processing efficiency in bilateral visual regions involved in the initial processing of the items. However, under conditions of high lexical competition (i.e., multiple completion word stems), the lexical processes, which are dominant in the left hemisphere, overshadow the visual priming supported by the left hemisphere. [ABSTRACT FROM AUTHOR]
- Published
- 2003
- Full Text
- View/download PDF
10. High Gamma Power Is Phase-Locked to Theta Oscillations in Human Neocortex.
- Author
-
Canolty, R. T., Edwards, E., Dalal, S. S., Soltani, M., Nagarajan, S. S., Kirsch, H. E., Berger, M. S., Barbaro, N. M., and Knight, R. T.
- Subjects
- *
NEOCORTEX , *BRAIN , *COGNITION , *CEREBRAL cortex , *OSCILLATIONS , *FLUCTUATIONS (Physics) , *AMPLITUDE modulation , *ROBUST control , *GAMMA (Electronic computer system) - Abstract
We observed robust coupling between the high- and low-frequency bands of ongoing electrical activity in the human brain. In particular, the phase of the low-frequency theta (4 to 8 hertz) rhythm modulates power in the high gamma (80 to 150 hertz) band of the electrocorticogram, with stronger modulation occurring at higher theta amplitudes. Furthermore, different behavioral tasks evoke distinct patterns of theta/high gamma coupling across the cortex. The results indicate that transient coupling between low- and high-frequency brain rhythms coordinates activity in distributed cortical areas, providing a mechanism for effective communication during cognitive processing in humans. [ABSTRACT FROM AUTHOR]
- Published
- 2006
- Full Text
- View/download PDF
Catalog
Discovery Service for Jio Institute Digital Library
For full access to our library's resources, please sign in.